Lamp housing Inspection sheet

32

Page

6 Lamp housing Inspection sheet

Study the instruction manual for the lamp housing before inspection.

For safe use of the lamp housing, we recommend performing the following inspection

periodically (every time you replace

the lamp bulb and at least every 6 months).

The table below identifies the check items to be observed. Put (X) if not applicable or (

)

if applicable.

If there is any (X) mark noted, immediately stop use of the product, and contact Olympus

for detailed inspections or replace the lamp housing.

If you detect an abnormality other than that listed below or with other Olympus product,

also stop the use of the product and contact Olympus for detailed inspections.

Note that the service, replacement and detailed inspections are charged after expiration of

the warranty period.

If you have any questions, please contact Olympus.

Check results (Date)

Check

items

/ / / /

1. More than 8 years have passed since original purchase or the total

power ON time has exceeded 20,000 hours.

2. Lamp does not light sometimes even though the power switch is set

to on.

3. Illumination flickers when you move the lamp cable or lamp housing.

4. Lamp cable is unusually hot to the touch.

5. Scorching or burning odor is produced during use.

6. Illumination still flickers after replacement with a new lamp bulb.

7. Deformation, backlash, or looseness, etc. when you assemble the

lamp housing.
(Impossibility of removing the top section of lamp housing when you
attempt to replace the lamp bulb, etc.)

8. Extreme discoloration of the lamp housing connection terminal or

lamp socket.
Uneven discoloration of the left and right sections of these parts.

9. Discoloration, deformation or cracking of the lamp housing.

10. Melting, crack, deformation or solidification of the lamp cable or a

wiring part.

11. Increased frequency of servicing compared to similar devices put

into use at the same time as the lamp housing.