Olympus VS120 User Manual
Page 35
Specifications
31
Page
■ Operating environment
Item
Specification
- Indoor
use
Altitude Max.
2000
meters
Operating temperature
15 to 28°C
Operating humidity
30 to 80% RH (no condensation)
Supply voltage fluctuations
±
10%
Overvoltage category
II (IEC60664)
Degree of contamination
2 (IEC60664)
■ Rated voltage/current
Before connecting each of the following pieces of equipment to the AC outlet, check the capacity of the AC
outlet to be used.
Computer
VS120-PC-E
100-120/200-240V ~
50/60Hz 10A/6A
Microscope controller
BX-UCB
100-120/220-240V ~
50/60Hz 3.5A/1.5A
Monitor
VS-MON24W
100-240V ~
50/60Hz 1.5A
Slide loader
VS120-LD100
100-120/220-240V ~
50/60Hz 0.9A/0.5A