beautypg.com

Transfer characteristics, Amplifier characteristics – National Instruments DAQCard-1200 User Manual

Page 71

background image

Appendix A

Specifications

DAQCard-1200 User Manual

A-2

© National Instruments Corporation

Maximum working voltage
(signal + common mode) ........................Negative input/AISENSE should

remain within ±6 V (bipolar) or
–6 to 2 V (unipolar) of AGND

Overvoltage protection ...........................±35 V powered on and ±25 V

powered off

Inputs protected ......................................

FIFO buffer size......................................1,024 samples

Data transfers ..........................................Interrupts, programmed I/O

Dither ......................................................Available

Transfer Characteristics

Relative accuracy....................................±0.5 LSB typ dithered,

±1.5 LSB max undithered

DNL ........................................................±1 LSB max

No missing codes....................................12 bits, guaranteed

Offset error

Pre-gain error after calibration ........10

µ

V max

Pre-gain error before calibration .....±20 mV max

Post-gain error after calibration.......1 mV max

Post-gain error before calibration ....±200 mV max

Gain error (relative to calibration reference)

After calibration...............................0.02% of reading max

Before calibration ............................±2% of reading max

Gain

1 with gain error

adjusted to 0 at gain = 1...................±0.8% of reading max

Amplifier Characteristics

Input impedance

Normal powered on .........................100 G

in parallel with 50 pF

Powered off .....................................4.7 k min

Overload ..........................................4.7 k min