Downloading iolan firmware, Calibrating i/o, Calibrating analog input – Perle Systems IOLAN SCS User Manual
Page 312: Calibrating voltage calibrating current
![background image](/manuals/180697/312/background.png)
Downloading IOLAN Firmware
312
IOLAN SDS/SCS/STS User’s Guide, Version 3.6
Downloading IOLAN Firmware
To upgrade the IOLAN firmware (software):
z
In DeviceManager, select
Tools
,
Advanced
,
Download Firmware to
IOLAN. You can browse to
the firmware location. Once the firmware download is complete, you will be prompted to reboot
the IOLAN. You can choose to reboot the IOLAN at another time by selecting
Tools
,
Reset
,
Reboot
IOLAN.
z
In WebManager, under the
Administration
option, select
Update Firmware
. Either browse to the
firmware file and then click the
Upload
button or configure the TFTP server and click the
Upload
button. Note: If you use the TFTP option, the specified TFTP server must be on the same
subnet as the IOLAN.
Upgrading the firmware does not affect the IOLAN’s configuration file or downloaded custom files.
Calibrating I/O
All I/O channels are factory calibrated and should not need recalibration during initial use. However
should calibration be required, you can recalibrate in DeviceManager or WebManager. In
DeviceManager, you calibrate the I/O channel(s) by selecting
Tools
,
I/O Channels
,
Calibrate
. In
WebManager, you calibrate the I/O channel(s) by selecting
I/O Channels
,
Calibrate
.
Calibrating Analog Input
To calibrate an Analog input channel, read the section that applies to the type of input you are
calibrating. Note that calibration will be done for the active channel configuration; for example, if
Channel A1 is set to voltage, you cannot calibrate it for current. The voltage range configured for this
channel will also dictate what is being calibrated. For example, if this channel is configured for a
range of +/-10V, calibrating this channel will calibrate all channels which are configured for +/-10V.
During the calibration process, you will be asked to apply the minimum and maximum configured
range value to the channel; for example, to calibrate for voltage +/- 10V, you will be prompted to first
apply -10V and then +10V to the channel.
Also, you cannot actively calibrate disabled channels (although, for Voltage, if you enable the
channel and then set it for a range that has already been calibrated for another channel, it will also be
calibrated).
Calibrating Voltage
When calibrating the IOLAN Analog input for voltage, you will need a calibration meter that is better
than .1% volts precision. When you calibrate one channel, all voltage channels are automatically
calibrated for that range; if another channel is set for a different range, you will need to calibrate that
channel separately, but all channels that use that range are also automatically calibrated.
Calibrating Current
When calibrating the IOLAN Analog input for current, you will need a calibration meter that is better
than .1% current precision. Each channel needs to be calibrated individually.