1316A-2 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 79 of 97

2. Type in the correct concentrations for the ticked CO2, Freon

gas.

The gas concentration must be the same as that stated on the

calibration gas bottle.

3. Connect the calibration-gas bottle as described in

section

4.3.4

.

4. Open the gas valve and adjust the flow so that a small surplus

flow exists in the side-stream tube.

5. Press the Send button to start the gain calibration.

After making the Span Calibration please make sure that the Mode is

set back into 5 and that the Zero calibration is set to “Auto”, as

described in

section 4.3.5

.

The monitor is now ready for making measurements.

Press Send.

The Process State In Progress is displayed in the Span Calibration

window (see

figure 4.30

) and the Data Status in the Status window is

changed to show Freon Span, see

figure 4.31

.

The 1316A-2 is calibrated with the selected gas at the concentration

keyed in.

Figure 4.30. Span Calibration window showing Freon calibration in

Progress.