2 the infrared transmission spectra, 3 the ndir module building blocks – LumaSense Technologies INNOVA 1316A-2 User Manual
Page 12
Chapter 1
_________________________________________________________________________________________
_____________________________________________________________________________
BE6021-11
1316A-2 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 12 of 97
Fig.1.2 Simple NDIR Gas Monitor
1.4.2 The Infrared Transmission Spectra.
The concentration of a subject gas in the sample is computed as a
function of the light intensity measured by the detector. The figure
below characterizes the band pass filters for Freon and CO
2
gases.
Filter
Centre
Wavelength
Band
Width
Centre
Wavelength
Band
Width
μm
μm
cm-1
cm-1
Freon
8.73
1.24
1145
163
Carbon
dioxide
4.425
0.07
2260
36
Fig.1.3 Infrared optical filters for Freon and Carbon Dioxide.
1.4.3 The NDIR Module Building blocks
Infrared Source
A current-regulated infrared (IR) source provides a photon stream.
Light is collimated and directed through the sample cell to the optical
block. The source is modulated at 1 Hz, with a duty cycle of 50%.