Figure 3.11: The Soft Reset Window.

Press the Send Button and the Instrument is Reset and ready for set-

up.

3.6 Operating Modes of the Multi Gas Monitor 1316A-2

The Multi Gas Monitor 1316A-2’s operations comprise four modes of

operation.

Start-up

Normal

Standby

System Fault

3.6.1 Start-up Operating Mode

After POR (power on / reset), the Multi Gas Monitor 1316A-2

performs the following:

Start-up and initialization.

Self-test.

Thermal warm-up. After a cold start or a Software Reset

bench command the warm-up time period is 35 seconds

maximum. After leaving standby mode the typical warm-

up time period is 20 seconds after completion of the first

zero.