6 operating modes of the multi gas monitor 1316a-2, 1 start-up operating mode – LumaSense Technologies INNOVA 1316A-2 User Manual
Page 49

Chapter 3
_________________________________________________________________________________________
_____________________________________________________________________________
BE6021-11
1316A-2 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 49 of 97
Figure 3.11: The Soft Reset Window.
Press the Send Button and the Instrument is Reset and ready for set-
up.
3.6 Operating Modes of the Multi Gas Monitor 1316A-2
The Multi Gas Monitor 1316A-2’s operations comprise four modes of
operation.
♦
Start-up
♦
Normal
♦
Standby
♦
System Fault
3.6.1 Start-up Operating Mode
After POR (power on / reset), the Multi Gas Monitor 1316A-2
performs the following:
♦
Start-up and initialization.
♦
Self-test.
♦
Thermal warm-up. After a cold start or a Software Reset
bench command the warm-up time period is 35 seconds
maximum. After leaving standby mode the typical warm-
up time period is 20 seconds after completion of the first
zero.