beautypg.com

4 the ndir temperature measurements – LumaSense Technologies INNOVA 1316A-2 User Manual

Page 14

background image

Chapter 1
_________________________________________________________________________________________

_____________________________________________________________________________

BE6021-11

1316A-2 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 14 of 97

System Board Assembly

The system board assembly is a surface-mount printed circuit board

that includes the digital and analogue electronic components and the

IR detector assembly. The IR detector assembly includes thermopile

IR detectors for the Freon, CO

2

and reference channels. Light passed

through the band pass filters is directed through the thermopile

window to a detector substrate. The substrate has a light-sensitive

coating, which produces a voltage that is proportional to light

intensity. A reference voltage is produced by the reference detector,

which provides an output that varies with light intensity and is not

affected by the sample gas. The detector block temperature

measurement facilitates compensation for IR filter and IR detector

temperature variances. The detector outputs are amplified and sent

through a multiplexer to an ADC converter. The microprocessor

samples the ADC converter output at 200 Hz and derives an

uncompensated digital value representing each gas concentration.

The data are subjected to a compensation routine that provides

adjustment for temperature and pressure variances. The digital value

representing each gas concentration is derived.

1.4.4 The NDIR Temperature Measurements.


The analyzer incorporates several thermistors for the measurement

of temperatures.

IR Detector/Filter Temperature :

A thermistor located in the optical assembly measures the IR

detector and IR filter temperature. It is used as follows:
“Gas law compensation” and detector response compensation during

Freon and CO

2

gas concentration measurement.

Span calibration (via the use of temperature compensated Freon and

CO

2

data).

An “out of range” IR detector/filter temperature condition causes

Freon and CO

2

data to be declared “data invalid.”

In the Module the IR Detector/Filter Temperature is monitored to set

the zero request bit as follows:
Zero Request if a ±5°C temperature change has occurred.
Ambient Temperature Out Of Range is set when the IR

Detector/Filter Temperature is not within its allowed range.