beautypg.com

LumaSense Technologies INNOVA 1316A-2 User Manual

Page 55

background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6021-11

1316A-2 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 55 of 97

Purpose

Zero calibration compensates for conditions that affect gas concentration

measurement. Examples of these conditions are as follows:
Changes in ambient temperature.
Short-term sample cell contamination caused by inadequate particle

filtering.
Long-term sample cell contamination caused by deposits on the sample

cell wall or windows.
Zero calibration execution time is typically 30 to 45 seconds, including a

purge time that depends on the pneumatics configuration and a fixed

calibration time of 20 seconds. The process will take longer under the

following conditions:

1.

Extended Purge time ( PT) is used.


Purge time ( PT) can be extended . Pull down the Bench Command menu

select Zero/O

2

Change the PT value).

Select the Zero/ O2 Span Calibration.
The following window is displayed.