The NDIR Module and the Electrochemical Sensors are to be specified

prior to purchase.

The Multi Gas Monitor INNOVA 1316A-2 is delivered as specified by

the Customer.

1.4 NDIR Gas Measurement


The Module VM0101A measures the following sample gas

concentrations: Carbon Dioxide (CO

2

) and Freon either with reference

to FREON1 or FREON2 (selectable from the BZ6013 Application

Software)
And the UA1372 is as an example given with an O

2

and NO sensor

installed.

1.4.1 Freon and CO2 Measurement

The NDIR measurement method uses fixed, non-scanning infrared

light frequencies to characterize Freon and CO

2

gas concentrations.

NDIR absorption profiles are the basis for measurement. The

concentration of a gas volume is a function of the quantity of gas

molecules in the sample. The absorption of infrared light increases

with the number of gas molecules in the light path. As the

concentration of infrared-absorbing gas increases, the transmission

of infrared light decreases. A basic NDIR measurement system

includes the following elements:

Sample

Delivery:

The portion of the gas is transported to the 1316A-2 where particulates are

removed before the gas is introduced to the NDIR module.

Infrared

Source:

An infrared light source produces a wide range of light at frequencies covering the

infrared band and extending into the visible spectrum.

Sample

Cell:

The gas to be measured is transported through a sample cell that allows infrared

light to pass through the sample.

Infrared

Filters:

An optical band pass filter is used to select a specific band of infrared light. The

filter wavelength is based on the gas to be measured. Separate infrared filters are

used in the measurement of Freon and CO

2

gas concentrations.

Infrared

Detector:

Infrared light not absorbed by the sample gas is transmitted to infrared detectors.

The detectors produce an output voltage that is proportional to the measured light.

See

figure 1.2