B.8 initiate:continuous command, B.9 initiate:continuous query, B.10 [source:]list:clear command – KEPCO KLP Series Developer's Guide, Rev 1 User Manual

Page 110: B.11 [source:]list:control command, B.12 [source:]list:control? query, B.13 [source:]list:count command, B.14 [source:]list:count? query, Init, B.10, Sour

background image

B-4

KLP 031113

B.8

INITiate:CONTinuous COMMAND

INIT:CONT

Syntax:

Short Form: INIT:CONT {ON | OFF} or {1 | 0} (1 = on, 0 = off)
Long Form: INITiate:CONTinuous {ON | OFF} or {1 | 0} (1 = on, 0 = off)
Default setting is OFF.

Description: INIT:CONT ON enables continuous triggers.; INIT:CONT OFF disables continuous triggers. If

INIT:CONT is OFF, then INIT:IMM arms the trigger system for a single trigger. If INIT:CONT is ON,
then the trigger system is continuously armed. INIT:CONT ON sets the WTG bit in the Questionable
Status register to 1 after each trigger event. Executing *RST command sets INIT:CONT to OFF. (See
example, Figure B-1.)

B.9

INITiate:CONTinuous QUERY

INIT:CONT?

Syntax:

Short Form: INIT:CONT?

Long Form: INITiate:CONTinuous?

Return Value: 1 or 0 (1 = on, 0 = off)

Description: Determines whether continuous triggers are enabled or disabled. Power supply returns value of

INIT:CONT flag: “1” = continuous triggers are enabled (INIT:CONT ON); “0” = continuous triggers dis-
abled (INIT:CONT OFF). (See example, Figure B-1.)

B.10 [SOURce:]LIST:CLEar COMMAND

LIST:CLE

Syntax:

Short Form: LIST:CLE

Long Form: LIST:CLEar>

Description: Clears all list entries by setting all pointers to 0. Also sets LIST:DIR to UP. Related Commands: All

LIST commands (See example, Figure B-2.)

B.11

[SOURce:]LIST:CONTrol COMMAND

LIST:CONT

Syntax:

Short Form: LIST:CONT {1 | 0} (1 = on, 0 = off)
Long Form: LIST:CONTrol>{1 | 0} (1 = on, 0 = off)

Description: If relay control set to LIST, allows user to energize (1 or on) or de-energize (0 or off) internal

relay using LIST commands. A LIST:CONT value (0 or 1) must be entered, regardless of relay con-
trol setting selected. Related Commands: LIST:CONT?.

B.12 [SOURce:]LIST:CONTrol? QUERY

LIST:CONT?

Syntax:

Short Form: LIST:CONT?

Long Form: LIST:CONTrol?

return value = 0 (off, de-energized) or 1 (on, energized)

Description: Identifies whether internal relay is energized (1) or de-energized (0). Related Commands:

LIST:CONT.

B.13 [SOURce:]LIST:COUNt COMMAND

LIST:COUN

Syntax:

Short Form: LIST:COUN Long Form: LIST:COUNt>

Description: Establishes how many times the list is executed. Allows user to establish how many times the list

is executed. The order (beginning to end or end to beginning) is determined by LIST:DIR. For
LIST:COUN 0, the unit will execute the sequence indefinitely until a VOLT:MODE FIXED command is
received. Related Commands: LIST:DIR. (See example, Figure B-2.)

B.14 [SOURce:]LIST:COUNt?

QUERY

LIST:COUN?

Syntax:

Short Form: LIST:COUNT?

Long Form: LIST:COUNt?

Return Value:

Description: Identifies how many times the list will be executed. Returns value set by LIST:COUN command.

(See example, Figure B-2.)