Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-16

Units


Set the unit of the Y-axis to dBm, dBmV, dBuV, Volts or Watts. Wherein, dBm,
dBmV and dBuV are for Log scale; Volts and Watts are for Linear scale. The default
is dBm.

Key Points:

The conversion relationships between units are as follows.



×

=

W

R

Volts

B

001

.

0

1

log

10

m

d

2

(2-4)





=

V

Volts

V

B

µ

µ

1

log

20

d

(2-5)

=

V

Volts

V

B

m

1

log

20

m

d

(2-6)

R

Volts

Watts

2

=

(2-7)

Wherein,

R

denotes the reference resistance.


Ref Offset


Assign an offset to the reference level to compensate for gains or losses generated
between the device under measurement and the analyzer.

Key Points:
 The change of this value changes both the reference level readout and the

amplitude readout of the marker; but does not impact the position of the
curve on the screen.

 You can modify this parameter using the numeric keys. For more details,

please refer to Parameter Setting”.