Freq offset – RIGOL DSA875 User Manual
Page 57

Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800 Series
2-5
Table 2-3 Stop Frequency
Parameter
Explanation
Default
7.5 GHz
Range*
0 Hz to 7.5 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Span > 0, step = Span/200
Span = 0, step = VBW/100
Min = 1 Hz
Direction Key Step CF step
Note: *The range is from 100 Hz to 7.5 GHz in non-zero span.
Freq Offset
You can set a frequency offset value to account for frequency conversions between
the device under test and the input terminal of the spectrum analyzer.
Key Points:
The change of this parameter only changes the display values of the center
frequency, start frequency and stop frequency; but does not impact any
hardware settings of the spectrum analyzer.
You can modify this parameter using the numeric keys, knob or direction keys.
For more details, please refer to “Parameter Setting”.
To eliminate an offset, you can perform Preset operation or set the frequency
offset to 0Hz.
Table 2-4 Frequency Offset
Parameter
Explanation
Default
0 Hz
Range
-100 GHz ~ 100 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
37.5 MHz
Direction Key Step CF Step