Sweep, Tracking generator (option), Trigger – RIGOL DSA875 User Manual
Page 224

RIGOL
User’s Guide for DSA800 Series
5-10
Sweep
Sweep
DSA815
DSA832
DSA875
Sweep
time
span ≥ 100 Hz 10 ms to 1500 s 1 ms to 3200 s 1 ms to 7500 s
zero span
20 μs to 1500 s 20 μs to 3200 s 20 μs to 7500 s
Sweep
time
uncertainty
span ≥ 100 Hz 5% (nom.)
zero span
(sweep time
setting
value > 1 ms)
5% (nom.)
Sweep mode
continuous, single
Tracking Generator (Option)
TG Output
DSA815
DSA832
DSA875
Frequency range
100 kHz to 1.5
GHz
100 kHz to 3.2
GHz
100 kHz to 7.5
GHz
Output level range
-20 dBm to 0
dBm
-40 dBm to 0 dBm
Output level resolution
1 dB
Output flatness
relative to 50 MHz
±
3 dB (nom.)
Trigger
Trigger
Trigger source
free run, video, external
External trigger level
5 V TTL level