beautypg.com

Full span, Zero span – RIGOL DSA875 User Manual

Page 62

background image

RIGOL

Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-10

For more details, please refer to Parameter Setting”.


Table 2-6 Span

Parameter

Explanation

Default

7.5 GHz

Range*

0 Hz to 7.5 GHz

Unit

GHz, MHz, kHz, Hz

Knob Step

Span/200, Min = 1 Hz

Direction Key Step in 1, 2, 5 sequence

Note: *0 Hz is available only in zero span.


Full Span


Set the span of the analyzer to the maximum.

Zero Span


Set the span of the analyzer to 0 Hz. Both the start and stop frequencies will equal
the center frequency and the horizontal axis will denote time. The analyzer
measures the time domain characteristics of the amplitude of the corresponding
frequency point on the input signal.

Key Points:
As opposed to Non-zero span, the screen shows the time domain characteristics of
the fixed frequency component of the signal in zero span mode.
The following functions are invalid in Zero span mode:
FREQ: Peak->CF and Signal Track;
SPAN: Zoom In and Zoom Out;
Marker->: Mkr->CF, Mkr->Step, Mkr->Start, Mkr->Stop, MkrΔ->CF and

MkrΔ->Span;

MarkerReadout: Frequency, Period and 1/ΔTime (valid in Delta marker

type);

TG: Power Sweep (only for DSA815-TG/DSA832-TG/DSA875-TG).

This manual is related to the following products: