Chapter 4 Troubleshooting&Message

RIGOL

User’s Guide for DSA800 Series

4-5

(2) Execution Error:

Indicate an execution error is detected by the Execution Control Block.
Possible reasons are:

 A parameter following the command header is evaluated by the

device as being outside of its legal input range.

 The command could not be properly responded to due to the current

device condition.

Message number: -299 to -200.

(3) Device Specific Error:

The command cannot be correctly executed due to the current settings of
the hardware and software of the device.

Message number: -399 to -300 (defined in SCPI standard) and 300 to
1000.

(4) Query Error:

Indicate a query error is generated when the output queue controller
detects a Message Exchange Protocol error (see IEEE488.2,6.1.10).
Possible reasons are (see IEEE488.2,6.5.7) :

 When trying to read the output queue, the queue has no data or is

being suspended.

 Data in the output queue has been lost.

Message number: -499 to -400.

3. Status Message:

Warn you that the analyzer is now in a certain or abnormal condition and
should be paid attention to. Status messages are always marked with .
They are displayed on the screen until the instrument exits the condition, the

Esc key is pressed or the *CLS command is sent. The status messages are

stored in the corresponding status register, and can be queried by
sending :Status or :SYSTEM:ERROR[:NEXT]?.
Message number: 200 to 299.