Chapter 2 Front Panel Operation

RIGOL

User’s Guide for DSA800 Series

2-75

Table 2-51 Average Number for TOI Measurement

Parameter

Description

Default

10

Range

1 to 1000

Unit

N/A

Knob Step

1

Direction Key Step

1


2. Avg Mode

Set the average operation mode to “Exp” or “Repeat” and the default is “Exp”.
 When “Exp” is selected, the result is the exponential average of the

current N (N is specified in Avg Num) measurement results.

 When “Repeat” is selected, the result is the arithmetic average of the

current N (N is specified in Avg Num) measurement results.


3. Span

This span which is the same with the span of the analyzer is the frequency
range for sweep. Modifying this parameter will also change the span of the
analyzer. You can use the numeric keys, knob or direction keys to modify this
parameter.

Table 2-52 Span for TOI Measurement

Parameter

Description

Default

2 MHz

Range

100 Hz to 7.5 GHz

Unit

GHz, MHz, kHz, Hz

Knob Step

TOI Distortion Span/100, the
minimum is 1 Hz

Direction Key Step

in 1-1.5-2-3-5-7.5 sequence