Chapter 2 Front Panel Operation

RIGOL

User’s Guide for DSA800 Series

2-7

Start

Execute a

sweep

Search for

peak and

mark it as

Marker

Active Marker

exists?

N

Set marker

frequency to

center freq

Y

Execute the

next sweep

Search for and mark the

frequency point (variation

< 3dB) near the Marker

Figure 2-1 Process of Signal Track


Key Points:
 When Signal Track is turned on, the ST (Signal Track) icon

is highlighted

in the status bar at the left of the screen.

 If an active marker currently exists, when Signal Track is enabled, the

instrument will search for and mark the point (with no more than 3 dB
variation in amplitude) near the marker as well as set the frequency of this
point as the center frequency to hold the signal at the center of the screen.

 If no marker is currently active, when Signal Track is enabled, the instrument

will activate Marker 1, execute a peak search automatically and set the
frequency of the current peak as the center frequency to hold the signal at the
center of the screen.

 In continuous sweep, the system tracks the signal continuously; in single

sweep, the instrument only performs a single signal track; in Zero Span,
Signal Track is invalid.