Chapter 4 Troubleshooting&Message
RIGOL
User’s Guide for DSA800 Series
4-19
359
Invalid function when trace is not enabled.
A marker could not be assigned to a trace when the trace is not enabled.
360
User preset failed.
The user preset has failed because the edition of the saved status data is
incorrect or the status data has been damaged. In this situation, the
system will execute Factory settings as the defaults.
400
Input signal power out of range.
Indicate that the input signal power is out of range. Working in this state
for a long time would cause damage to the device. Please reduce the
input power.
401
1st LO unlock.
The first LO is unlocked. Please report the error to your
RIGOL sales or
technical support.
402
2nd LO unlock.
The second LO is unlocked. Please report the error to your
RIGOL sales
or technical support.
403
Track LO unlock.
Track LO is unlocked. Please report the error to your
RIGOL sales or
technical support.
412
DA overrange due to volume adjustment in FM.
413
Intermediate frequency signal out of range.
420
Option not installed.
The desired operation cannot be performed because the option required
is not installed.
460
No peak found.
No signal peak that meets the criteria was found under Peak Search.
461
No fundamental wave found.
Indicate that no fundamental wave greater than 50dBm is found.
462
No two-tone signal found.
463
Invalid function as cursor is not enabled.
The cursor cannot be set to the reference level unless it is enabled.
464
Menus under Storage can not be set to Userkey.
465
Invalid function.
470
Calibration memory lost.
Indicate that the calibration data before delivery or last valid calibration
data has been lost. Please report the error to your
RIGOL sales or
technical support team.