RIGOL DSA875 User Manual
Page 122
![background image](https://www.manualsdir.com/files/859146/content/doc122.png)
RIGOL
Chapter 2 Front Panel Operation
User’s Guide for DSA800 Series
2-70
Table 2-46 Noise Bandwidth for C/N Ratio Measurement
Parameter
Description
Default
2 MHz
Range
33 Hz to 2.5 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Noise BW/100, the minimum is 1 Hz
Direction Key Step
in 1-1.5-2-3-5-7.5 sequence
5. Carrier BW
Set the bandwidth of the carrier to be measured.
The carrier bandwidth is related to the noise bandwidth and the range
available is from noise bandwidth/20 to noise bandwidth×20.
You can use the numeric keys, knob or direction keys to modify this
parameter.
Table 2-47 Carrier Bandwidth for C/N Ratio Measurement
Parameter
Description
Default
2 MHz
Range
33 Hz to 2.5 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Carrier Noise/100, the minimum is 1 Hz
Direction Key Step
in 1-1.5-2-3-5-7.5 sequence