beautypg.com

RIGOL DSA875 User Manual

Page 112

background image

RIGOL

Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-60

Table 2-35 Average Number for Chan Power Measurement

Parameter

Description

Default

10

Range

1 to 1000

Unit

N/A

Knob Step

1

Direction Key Step

1


2. Avg Mode

Set the average operation mode to “Exp” or “Repeat” and the default is “Exp”.
 When “Exp” is selected, the result is the exponential average of the

current N (N is specified in Avg Num) measurement results.

 When “Repeat” is selected, the result is the arithmetic average of the

current N (N is specified in Avg Num) measurement results.


3. Integ BW

Set the frequency width of the channel to be tested and the power of the
channel is the power integral within this bandwidth. You can use the numeric
keys, knob or direction keys to modify this parameter.

Table 2-36 Integration Bandwidth for Chan Power Measurement

Parameter

Description

Default

2 MHz

Range

100 Hz to 7.5 GHz

Unit

GHz, MHz, kHz, Hz

Knob Step

Integration BW/100, the minimum is 1 Hz

Direction Key Step

in 1-1.5-2-3-5-7.5 sequence

4. CH Pwr Span

Set the frequency range of the channel. This span which is the same with the
span of the analyzer is the frequency range of the sweep. Modifying this
parameter will change the span of the analyzer.
 The channel power span is related to the integration bandwidth and the

range available is from integration bandwidth to integration
bandwidth×20.

 You can use the numeric keys, knob or direction keys to modify this

parameter.

This manual is related to the following products: