Trigger Input
When signal that meets the current trigger type is input on the specified data line,
the specified source under control will turn on the output, turn off the output or
toggle the output state according to the setting in output response.
Press
Trigger Trig to select "In" and the trigger input setting interface is
opened, as shown in Figure 2-6.
1. Data Line
Press Data Line to select D0, D1, D2 or D3 and the data line currently selected
will be highlighted on the screen. Users can set the trigger conditions of the four
data lines respectively.
2. Source under Control
Press Ctrled Src to select one or more of CH1, CH2 and CH3 as the source
under control.
3. Trigger Type
Press Trig Type to select to trigger on the rising edge, falling edge, high level or
low level of the input signal.
Tip
For the input signal, high level is from 2.5V to 3.3V, low level is from 0V to 0.8V
and noise tolerance is 0.4V.
4. Output Response
Press Outp Resp to set the output response type to "Output On", "Output Off"