beautypg.com

Trigger input – RIGOL DP832 User Manual

Page 88

background image

RIGOL

Chapter 2 Front Panel Operations

2-38

DP800 User’s Guide

Figure 2-6 Trigger Setting Interface

Trigger Input

When signal that meets the current trigger type is input on the specified data line,

the specified source under control will turn on the output, turn off the output or

toggle the output state according to the setting in output response.


Press

TriggerTrig to select "In" and the trigger input setting interface is

opened, as shown in Figure 2-6.

1. Data Line

Press Data Line to select D0, D1, D2 or D3 and the data line currently selected

will be highlighted on the screen. Users can set the trigger conditions of the four

data lines respectively.

2. Source under Control

Press Ctrled Src to select one or more of CH1, CH2 and CH3 as the source

under control.

3. Trigger Type

Press Trig Type to select to trigger on the rising edge, falling edge, high level or

low level of the input signal.

Tip

For the input signal, high level is from 2.5V to 3.3V, low level is from 0V to 0.8V

and noise tolerance is 0.4V.

4. Output Response

Press Outp Resp to set the output response type to "Output On", "Output Off"