beautypg.com

To set the delay parameters, To set the number of output groups, To set the delay parameters -26 – RIGOL DP832 User Manual

Page 76

background image

RIGOL

Chapter 2 Front Panel Operations

2-26

DP800 User’s Guide

To Set the Delay Parameters

Press TimerDelay Set to enter the delay parameter setting interface as shown
in the figure below. The channel currently selected is displayed in the status bar.

Press the channel selection keys at the front panel to switch the channel currently

selected. This interface provides delay parameter preview. Users can view the values

on the current page of the delay parameter list (high level indicates turning on the

output and low level indicates turning off the output).

Figure 2-4 Delay Parameter Setting Interface

You can set the delay parameters in the delay parameter setting interface following

the steps below.

1. To Set the Number of Output Groups

2. To Set the Number of Cycles

3. To Set the End State

4. To Edit the Delay Parameters

5. To Set the Stop Condition

6. To Save and Read the Delay File

To Set the Number of Output Groups

The number of output groups refers to the number of times that the instrument turns

on or off the output according to the preset state. Press Groups; use the numeric

keyboard or the left/right direction key and knob to input the desired value. The

range is from 1 to 2048.

Delay Parameter Preview Channel Currently Selected Delay Parameter List