beautypg.com

Edit the timer parameters – RIGOL DP832 User Manual

Page 73

background image

Chapter 2 Front Panel Operations

RIGOL

DP800 User’s Guide

2-23

3. Edit the timer parameters

For different templets, the timer parameters to be set are different as shown in

Table 2-1.

Table 2-1 Timer parameters (templet)

Templet Parameter
Sine

Max Value, Min Value, Points, Interval, Inverted

Pulse

Hi Level, Lo Level, Width, Period, Inverted

Ramp

Max Value, Min Value, Points, Interval, Symmetry, Inverted

Stair Up

Max Value, Min Value, Points, Interval

Stair Dn

Max Value, Min Value, Points, Interval

StairUpDn Max Value, Min Value, Points, Interval
Exp Rise

Max Value, Min Value, Points, Interval, Rise Index

Exp Fall

Max Value, Min Value, Points, Interval, Fall Index

 Max Value

Set the maximum voltage or current of the templet currently selected. The

range is related to the channel currently selected. It should be greater than

or equal to the minimum value currently set.

 Min Value

Set the minimum voltage or current of the templet currently selected. The

range is related to the channel currently selected. It should be lower than or

equal to the maximum value currently set.

 Points

Set the total number of points (namely the number of output groups in the

timer parameters) of the templet currently selected. The range is from 10 to

2048.

When the total number of points (denoted by P) and the current number of

output groups (denoted by G) are different, the number of output groups

will change to P automatically after output waveform is created using the

templet.

 Interval

Set the time interval between the points (the time required for the

instrument to output each group of timer parameters) of the templet

currently selected. The range is from 1s to 99999s.

 Inverted

When the templet currently selected is Sine, Pulse or Ramp, if invert is

enabled, the instrument will first turn the preset waveform upside down

and then create the output waveform.