RIGOL
5-4
DP800 User’s Guide
Mechanical
Dimensions
239mm(W) x 157mm(H) x 418mm(D)
Weight
DP832: 10.5kg
DP831: 9.75kg
DP821: 10.0kg
DP811: 10.3kg
Power
AC Input
(50Hz-60Hz)
100Vac±10%, 115Vac±10%, 230Vac±10% (maximum 250Vac)
Maximum Power
DP832: 521VA
DP831: 416VA
DP821: 450VA
DP811: 503VA
I/O
USB Device
1
USB Host
1
LAN
1 (Option)
RS232
1 (Option)
Digital IO
1 (Option)
USB-GPIB
1 (Option, extend a GPIB interface using the USB-GPIB interface
converter)
Rear Output Interface 1 (only for DP811)
Environment
Cooling Method
Fan Cooling
Working Temperature 0℃ to 40℃
Storage Temperature
-40℃ to 70℃
Humidity
5% to 80% relative humidity
Altitude
Below 1500m
Note
[1]
: The accuracy param eters are acquired via calibration under 25℃ after 1-hour warm-up.
Note
[2]
: The maximum time required for the output to change accordingly after receiving the
APPLy and SOURce commands.
Note
[3]
: The variation of the output within 8 hours after 30-minute warm-up when the load circuit
and environment temperature are constant.