beautypg.com

RIGOL DP832 User Manual

Page 69

background image

Chapter 2 Front Panel Operations

RIGOL

DP800 User’s Guide

2-19

Sine

The Sine waveform is as shown in the figure below. The instrument

determines the Sine amplitude according to the maximum and minimum

currently set and determines the Sine period according to the total number

of points (denoted by P) and the time interval currently set, thus

determining the Sine waveform. When creating parameters, the instrument

draws P values from the preset Sine waveform according to the current

time interval.

T

Pulse

The Pulse waveform is as shown in the figure below. The timer parameters

created from Pulse waveform only contain two groups of data.

The first group: the amplitude (voltage or current) is determined by the

high level set; the time equals the pulse width (t) currently set.

The second group: the amplitude (voltage or current) is determined by the

low level set; the time equals the period currently set minus the pulse width

currently set (T-t).

T

t

Tip

When Pulse is selected, if you want to output more than 2 groups of timer

parameters, you can output the timer parameters created from the Pulse

templet repeatedly by increasing the number of cycles.