Power-on Inspection

Press the power switch at the front panel, the instrument starts and executes

self-test. If the instrument passes the self-test, the welcome interface will be

displayed; otherwise, the corresponding self-test failure information (including the

top board, bottom board, fan and temperature) will be displayed.

Tip

When powering on the instrument after powering off it, make sure that the time

interval between the two operations is greater than 5s.

To Replace the Fuse

The fuse specification is related to the instrument model and actual input voltage, as

shown in the table below. You can also refer to the "Input Power Requirements" at

the rear panel of the instrument.

Table 1-3 Fuse specifications

Input Voltage

Fuse Specification

DP832/DP811
100Vac/115Vac

T6.3A

230Vac

T3.15A

DP831/DP821
100Vac/115Vac

T5A

230Vac

T2.5A

To replace the fuse, follow the steps below.

1. Turn off the instrument and remove the power cord.

2. Insert a small straight screwdriver into the slot at the power socket and prize out

the fuse seat gently.

Fuse Seat

Fuse