RIGOL DP832A User Manual
Page 58
RIGOL
Chapter 2 Front Panel Operations
2-4
DP800A User’s Guide
WARNING
To avoid electric shock, please turn on the output switch after the
output terminals are correctly connected.
CAUTION
When the fan stops, the channel switch can not be turned on;
otherwise, “The fan stops, stop the output!” will be displayed.
8. Check the output mode
In constant voltage output mode, the output mode displayed should be “CV”; if
“CC” is displayed, you can increase the current setting value properly and the
power supply will switch to CV mode automatically.
Besides, DP811A also provides output interface at the rear panel for constant voltage
output. As shown in the figure below, the load is connected to the output interface at
the rear panel for constant via a connecting terminal.
Connection Steps:
1. Connect the load to the corresponding position of the connecting terminal
correctly according to the figure above. Note the polarity when connecting.
2. Insert the connecting terminal to the output interface at the rear panel of the
instrument correctly according to the figure above. Note the corresponding
relationship between the connecting terminal and the output interface when
inserting.
Rear Output Interface
Connecting Terminal