Chapter 2 Front Panel Operations

RIGOL

DP800A User’s Guide

2-69

To Recall the user-defined setting

In the interface as shown in Figure 2-14, press Preset Key to select the specified
user-defined setting (take user-defined setting 1 as an example) to enter the
user-defined setting interface (the current user-defined setting is displayed in the
interface) as shown in Figure 2-15. At this point, pressing Preset at the front panel
will recall the corresponding user-defined setting.

Figure 2-15 User-defined Setting Interface

You can set the user-defined setting currently selected according to the steps below

in the interface as shown in Figure 2-15.

1. Use the left/right direction key and knob to edit the name of the current

user-defined setting (the default names of the 4 user-defined settings are user1,

user2, user3 and user4 respectively).

2. Press the channel selection keys at the front panel to select the desired channel

(the channel currently selected is highlighted).

3. Press Voltage (Current) and use the numeric keyboard or the left/right

direction key and knob to set the voltage (current) of the channel currently

selected.

4. Press OVP (OCP) and use the numeric keyboard or the left/right direction key

and knob to set the OVP (OCP) value of the channel currently selected. Press

OVP (OCP) repeatedly to enable or disable the OVP (OCP) function of the

channel currently selected.

5. Press Track repeatedly to enable or disable the track function of the channel

currently selected.