Open Collector Output Connections
Output connections are made to two terminals labeled OUTPUT. Connect to an input
device such as alarm indicator as shown in Figure 6, or drive a relay as shown in
Figure 7.
WARNING
To avoid damaging the PD6800-0L1’s amplifying components, use
care not to wire incorrectly or exceed output ratings. A diode, such
as 1N4000 series, will provide protection from relay transients.
Figure 6. Connection to Device with Internal Pull-Up
Figure 7. Output Connections
SAFE-TOUCH
BUTTONS
BACKLIGHT
POWER
UNLOCK
LOCK
LOOP
9-36 VDC
+ - + -
SIGNAL
BACKLIGHT
+ - + -
OUTPUT
RESET
+
Alarm Recorder
or Annunciator
(e.g. PD8-158)
-
SAFE-TOUCH
BUTTONS
BACKLIGHT
POWER
UNLOCK
LOCK
LOOP
9-36 VDC
+ - + -
SIGNAL
BACKLIGHT
+ - + -
OUTPUT
RESET
5-30 VDC
150 mA MAX