Electrical specifications – Sealey TA304 User Manual
Page 6
Note: Infrared measurement considerations.
Holding the laser unit in your hand, direct the laser/Ir sensor toward the
object whose temperature is to be measured bearing in mind the field of
view requirements shown in fig.12. the meter automatically
compensates for temperature deviations due to the ambient
temperature. If a higher temperature is to be measured after taking a
low temperature measurement, it is advisable to wait several minutes
between the measurements for the infrared sensor to cool. Keep in mind
that it may take up to 30 minutes to adjust to wide ambient temperature
differences.
INFRARED TEMPERATURE PROBE SPECIFICATION
response rate
0.5 seconds approx.
operating temperature
32ºf to 122ºf (0ºc to 50ºc)
operating Humidity
Max 80% rH.
Power supply
9V battery (PP3)
Weight
182g
size
164 x 50 x 40mm
Ir thermometer range
-30 to 550ºc (-22ºf to 1022ºf)
Ir thermometer output
1mV = 1ºc or 1ºf
Ir thermometer
accuracy
± 2% of reading or 4ºf whichever is greater. note:
accuracy is specified for the following ambient
temperature range: 64 to 82ºf (18 to 28ºc)
Emissivity
0.95 fixed value
field of view
D/s Approx. 8.1 (D = distance & s = spot) Has
90% encircled energy at focal point.
Laser Power
Less than 1mW
spectral response
6 to 14 µm (wavelength)
GENERAL INSTRUMENT SPECIFICATION
Instrument complies with
IEc 1010-1 En61010-1
Insulation
class 2, Double Insulation
overvoltage category
cAtIII1000V/cAtIV600V
Display
4000 counts LcD display with
function indication
Polarity
Automatic, (-) negative polarity indication.
overrange
"oL" marks indication
Low battery indication
"BAt" is displayed when the battery
voltage drops below the operating level
Measurement rate
2 times per second, nominal
Auto power off
Meter automatically shuts down after
approx. 30 seconds of inactivity
operating environment
0ºc to 50ºc (32ºf to 122ºf) at < 70%
relative humidity.
storage temperature
-20ºc to 60ºc (-4ºf to 140ºf) at < 80%
relative humidity.
for inside use, max height
2000m
Pollution degree
2
Power
one 9V battery (PP3)
Dimensions
182 x 82 x 55mm
Weight
360g
RPM (tach)
Range
Resolution
Accuracy
rPM4
600 ~ 4000 rPM 1 rPM
±2% of rdg ± 4dgts
1000 ~ 12000
rPM. (x10 rPM)
10rPM
rPM 2/DIs
300 ~ 4000 rPM 1 rPM
1000 ~ 6000
rPM. (x10 rPM)
10rPM
Effect reading: >600rPM
6. ELECTRICAL SPECIFICATIONS
Electrical Specification (ACCURACY).
Accuracy is given as ± ([% of reading] + [number of least significant
digits]) at 18ºc to 28ºc (65ºf to 83ºf), with relative humudity up to 70%.
DWELL ANGLE
Cylinder
Range
Resolution
Accuracy
4cYL
0 ~ 90.0º
0.1º
±2% of rdg ± 4dgts
5cYL
0 ~ 72.0º
6cYL
0 ~ 60.0º
8cYL
0 ~ 45.0º
DC VOLTAGE
Range
Resolution
Accuracy
400.0mV
0.1mV
±0.5% of rdg ± 3dgts
4.000V
1mV
±1.5% of rdg ± 2dgts
40.00V
10mV
400.0V
100mV
600V
1V
±1.8% of rdg ± 2dgts
Input Impedance: 10MΩ
AC VOLTAGE
Range
Resolution
Accuracy
400.0mV
0.1mV
±1.5% of rdg ± 5dgts
4.000V
1mV
±1.0% of rdg ± 3dgts
40.00V
10mV
±1.5% of rdg ± 3dgts
400.0V
100mV
600V
1V
±2.0% of rdg ± 4dgts
Input Impedance: 10MΩ / Frequency Range 50 to 60Hz
DC CURRENT
Range
Resolution
Accuracy
40.00mA
10uA
±1.5% of rdg ± 3dgts
400.0mA
100uA
20A
10mA
±2.5% of rdg ± 5dgts
AC CURRENT
Range
Resolution
Accuracy
40.00mA
10uA
±1.8% of rdg ± 5dgts
400.0mA
100uA
20A
10mA
±3.0% of rdg ± 7dgts
overload Protection: 0.5A / 250V and 20A / 250V fuse
Maximum Input: 400mA Dc or 400mA Ac rMs on uA / mA ranges,
20A Dc or Ac rMs on 20A range.
overload Protection: 0.5A / 250V and 20A / 250V fuse
frequency range: 50 to 60Hz
Maximum Input: 400mA Dc or 400mA Ac rMs on uA / mA ranges,
20A Dc or Ac rMs on 20A range.
RESISTANCE
Range
Resolution
Accuracy
400.0Ω
0.1Ω
±1.2% of rdg ± 4dgts
4.000kΩ
1Ω
±1.0% of rdg ± 2dgts
40.00kΩ
10Ω
400.0kΩ
100Ω
±1.2% of rdg ± 2dgts
4.000MΩ
1kΩ
40.00MΩ
10kΩ
±2.0% of rdg ± 3dgts
tA304 Issue:4(I) - 17/06/14
Original Language Version
© Jack sealey Limited