Error messages


Fig. 1

ERR19 (Excess Sensor
Current)
- Please replace Sensor!

The current drawn by the SmartSensor is too high and cannot be controlled
to the setpoint. Possibly liquid in the SmartSensor, or handset electronics is
defective.

REMEDY: Replace the SmartSensor. Replace the handset electronics.



>>>> ALLGEMEIN

WARNING:
SmartSensor lifetime is
nearly over!
Please use new P1:P
SmartSensor!

The SmartSensor has been operated for over 300 operating hours and is
approaching the end of its service life.

REMEDY: Replace the SmartSensor as soon as possible.

NOT READY !
SmartSensor lifetime is
over!
Please use new P1:P
SmartSensor!

The SmartSensor has reached its service life limit of 320 hours and must no
longer be used.

REMEDY: Replace the SmartSensor.