Sniffer probe extension
Part No.: 23004
The extension must not be used without installed tube filter. When changing the filter, avoid
touching the front part of the capillary.
The flexible probe extension enables you to search for leaks also at inaccessible places.
The detection limit is not influenced, the response time is increased to approx. 1.2 s.
The flexible stainless steel capillary can be bent if you need to, however, avoid sharp radii.
To use the extension, remove the standard probe cap from the measuring probe, and manually tighten the
extension. First, make sure that the front filter (Part No.: 23002) is present on the detector head.
The inner diameter of the capillary is 0.25 mm. The tube filter, placed on the front protects against
contamination from grease, etc. If the filter is dirty, the leak detector outputs an error message. In this case,
replace the tube filter (Part No.: 23005).
Screw cap
Al - reinforced
Flexible Al - tube
Plastic coated
Tube filter
Capillary
tube
Do not twist Flextube