Leads extender, Sf6 leak detector calibrator – Ion Science SF6 P1:p User Manual
Page 33
SF6 P1:p MANUAL
Ion Science Ltd
32 of 41
Unrivalled Detection. www.ionscience.com
Accessories
Length
300 mm, 500 mm, custom lengths upon request
Outside diameter
6 mm aluminium tube, plastic coated with inside
capillary
Response time
(t90)
/ recovery time
(t10)
1.0 s / 1.0 s (total time, including P1 SmartSensor)
Leads extender
Part No.: 23006 (5 meters)
Part No.: 23007 (10 meters)
The 5/10-m long extender is simply connected between the handset and the base unit. It increases the
operating range of the leak detector handset. Detection limit, response time and recovery time are not
influenced because the detector is built-into the sensor tip.
CAUTION: Make sure that the leak detector
is switched off before you use the leads extender. Only then
may the plug-in connections between the handset and base unit be disconnected, respectively reconnected.
SF6 Leak Detector Calibrator
Part No.: A-21520
The portable leak calibrator is filled with gas, sufficient for approximately 1000 calibrations. The integrated
sinter leak specified for long-term stability makes it possible to test and calibrate all SF
6
gas leak detectors in
the range between 10E-6 and 10E-7 ml/s.
The calibrator comes with an international calibration certificate, traceable to DKD (German Calibration
Service) standards. Multiple point calibration is also available as option.