Filter replacement
Parts overview
The ISM front filter optimises the replacement process for the suction filter as well as the service life of the
SmartSensor, which essentially depends on the amount contamination from the ambient air flowing into the
ionisation chamber.
Should a filter replacement become necessary then the leak detector should be switched off before un-
screwing the probe’s cap and removing the filter. The new filter element is then pushed over the detector
head, and the probe’s cap can be again tightened manually. When doing so, ensure cleanness.
Especially ensure that no particles are trapped between filter and detector head and make sure that all
components are firmly screwed together providing a tight seal.
P1:p Front filter
Part No.: 23002
P1:p Long cap (25mm)
Part No..: 23003
P1:p Probe extender,
flexible, 300/500 mm
Part No.: 23004
P1:p SmartSensor
Part No. 23001
P1:p Hose filter
Part No.: 23005