Setting up the operating parameters
This seal needs to be maintained during the entire running time of the automatic leak test. After the
automatic leak test has been completed, the result is displayed and the operator requested to remove the
seal.
This automatic test serves the purpose of identifying the presence of any leaks within the pneumatic system
of the SmartSensor. Currently a routine check of this kind, for example after having replaced the
SmartSensor is not recommended, since new, respectively refurbished SmartSensors are meticulously
checked before being delivered. Perform this test in coordination with your service partner or when receiving
upon starting the system a warning as to an operating vacuum which is too low.
[ OFF: SENSOR ] : Shuts down the sensor.
[ OFF: POWER ] : Switches the instrument off. When operated briefly indication 04 (Low Bat Warn) must
respond.
While running the diagnosis function, the LEDs in the handset as well as the analogue gauge are cyclically
driven so that the operation of these components can be checked visually.
Printing out all technical data
[ TECH. DATA READOUT ] / [ PRINT TECH. DATA ]
New to software version V3.06 is the capability of being able to transfer besides the technical data of base
unit, handset and SmartSensor also the calibration data as well as all operator specific settings to a printer or
a computer. Through the buttons [ CALIBRATION DATA ], [ USER SETTINGS ] and [TECHNICAL DATA ]
all required groups of data can be marked for printing out or removed from the printing out process. The
shortcut buttons [ ALL ] and [ NONE ] mark all, respectively no groups of data for printing out. The button [
GO ] starts the printing process.
As to how connect to a printer or computer, see Chapter 0, Handshake .
The supervisor level
(SUPERVISOR LEVEL)
The supervisor level ensures administration of the passwords as well as the settings for the limits and the
entries which are mandatory. The supervisor level is accessed from the main menu through the button [
OPEN SUPERVISOR LEVEL ] with subsequent entry of the supervisor password.
This level offers the following functions:
Making/deleting worker identification mandatory
[ OPEN SUPERVISOR LEVEL] / [ OPERATOR ID MODE ]
With worker identification having been made mandatory, the worker will then have to identify himself before
saving any measured values. If he has not done this at the point of time of saving the first measured value,
he will then be requested to do so. Without having entered an identification, no measured value can be
saved. The same applies to opening a new test specimen data set in the memory.
To make worker identification mandatory, touch the button [ REQUIRED ]. A check mark in the button
confirms the activation.
In the case worker identification has not been mandatory, the worker identification can be entered but is not
demanded from the operator.
To disable mandatory worker identification touch the button [ NOT REQUIRED ]. A check mark in the button
confirms the deactivation.
Making/deleting test specimen identification mandatory
[ OPEN SUPERVISOR LEVEL] / [ PRODUCT ID MODE ]
With test specimen identification having been made mandatory, the test specimen must be identified before
saving any measurement data by entering a serial or batch number, for example. If this is not done this at the
point of time of saving the first measured value, the operator then be requested to do so. Without having
entered an identification, no measured value can be saved. The same applies to opening a new test
specimen data set in the memory.
To make identification mandatory, touch the button [ REQUIRED ]. A check mark in the button confirms the
activation.