List of Operating Parameters
Parameter
Settings / Function
P45.2 ALARM
Encoder monitoring X2
No monitoring
ALARM OFF
Monitor the frequency
FREQUENCY
Monitor contamination
CONTAMINAT.
Contamination + frequency FRQ. + CONT.
P50 RS232
Baud rate
110 / 150 / 300 / 600 / 1200 /
2400 / 4800 / 9600 / 19 200 /
38 400 baud
P51 RS232
Additional blank lines
during data output
BK LINE 1
0
£ P51 £ÿ99
Default setting: 1
P62 A1
Trigger limit 1
P63 A2
Trigger limit 2
P79 PRESET
Value for datum point
Enter numerical value for datum setting over
switching input or with ENT key
P80 ENT-CL
Set display
No zero reset/Set with
CL/ENT
CL-ENT OFF
Zero reset with CL
setting disabled with ENT
CL......ON
Zero reset with CL and set with
ENT to value selected in P79
CL-ENT ON
Parameter
Settings / Function
P82 DISPL.ON
Message after switch-on
ENT...CL message
MESSAGE ON
No message
MESSAGE OFF
P85 EXT.REF
External REF
REF over D-sub port EXT
EXT.REF ON
No REF over D-sub port EXT EXT.REF OFF
P86 MOD
Print via MOD key
PRINT via MOD disabled
PRINT OFF
PRINT via MOD enabled
PRINT ON
P98 LANGUA.
Conversational language
German
LANGUAGE D
English
LANGUAGE GB
French
LANGUAGE F
Italian
LANGUAGE I
Dutch
LANGUAGE NL
Spanish
LANGUAGE E
Danish
LANGUAGE DK
Swedish
LANGUAGE S
Finnish
LANGUAGE FI
Czech
LANGUAGE CZ
Polish
LANGUAGE PL
Hungarian
LANGUAGE H
Portuguese
LANGUAGE P