List of operating parameters – HEIDENHAIN ND 231 B v.2 User Manual
Page 19
19
Parameter
Settings/ Function
P40.1 COMP.
Select encoder compensation X1
No compensation
COMP. OFF
Multipoint: up to 64
compensation points
COMP. MULTI
Linear compensation
COMP. LIN
1)
P40.2 COMP.
Select encoder compensation X2
No compensation
COMP. OFF
Multipoint: up to 64
compensation points
COMP. MULTI
Linear compensation
COMP. LIN
P41.1 L.COMP.
Linear error compensation X1
– 99 999.9 < P41 < + 99 999.9 [µm/m]
Default setting: 0
Example: Find input value for P41
Displayed measuring length ............. L
d
= 620.000 mm
Actual length (measured, e.g.
with the VM 101 comparator
system from HEIDENHAIN) .............. L
a
= 619.877 mm
Length difference .............................
DL = L
a
– L
d
= – 123 µm
Compensation factor k (= P41):
k =
DL / L
d
= – 123 µm / 0.62 m ....... k = – 198.4 [µm/m]
P41.2 L.COMP. Linear error compensation X2
– 99 999.9 < P41 < + 99 999.9 [µm/m]
Default setting: 0
Parameter
Settings/ Function
P30.2 DIR
Counting direction X2
Positive counting direction with
positive direction of traverse DIRECT. POS
Negative counting direction with
positive direction of traverse DIRECT. NEG
P31.1 S. PER.
Signal period of encoder X1
0.000 000 01 < P31 < 99 999.9999
Default setting: 10 µm
P31.2 S. PER.
Signal period of encoder X2
0.000 000 01 < P31 < 99 999.9999
Default setting: 10 µm
P33.1 STEP
Counting mode X1
0-1-2-3-4-5-6-7-8-9
COUNT 0-1
0-2-4-6-8
COUNT 0-2
0-5
COUNT 0-5
P33.2 STEP
Counting mode X2
0-1-2-3-4-5-6-7-8-9
COUNT 0-1
0-2-4-6-8
COUNT 0-2
0-5
COUNT 0-5
P38.1 DEC.
Decimal places X1
1)
1 / 2 / 3 / 4 / 5 / 6
(up to 8 for inch display)
P38.2 DEC.
Decimal places X2
1)
1 / 2 / 3 / 4 / 5 / 6
(up to 8 for inch display)
List of Operating Parameters
1)
Depends on signal period (P31) and unit of measure (P01)