12

Error Messages

Display

Problem

RS232 FAST

Command for measured value

output followed too quickly by

another.

1)

SIGNAL X1

Encoder signal is too weak.

The scale may be contaminated.

1)

DSR.MISSING

The connected device has not sent

a DSR signal.

1)

REF. ERR. X1

The spacing of the reference marks

as defined in P43 is not the actual

spacing.

1)

FORMAT ERR.

Data format, baud rate, etc. do not

match.

1)

FRQ.ERR. X1

Input frequency too high for encoder

input. Traversing speed may be too

fast.

1)

MEMORY ERR.

Checksum error: Check the datum,

operating parameters and compen-

sation values for multipoint axis

error compensation. If this error

recurs, contact your service

agency!

1)

These errors are important for the attached device. The

error signal (pin 19) at D-sub connection EXT is active.

Other error displays
If “OVERFLOW“ appears, the measured value is too large or
too small:
ä Set a new datum.

or

ä Move the axis back.

If

all sorting signals light up, the upper sorting limit is smaller

than the lower limit:

ä Change operating parameters P18 and/or P19.
To clear error messages:

When you have removed the cause of the error:

ä Clear the error message with the CL key.

Error

Messages

Display

Problem

REC. ERROR

Error during reception of parame-

ter and compensation value lists.