Shut down interface – Super Systems MGA6010 User Manual
Page 43
MGA 6010 Operations Manual
Super Systems Inc.
Page 43 of 54
The Valve Setup window contains the following items:
A – Valve Selection Drop-Down List. Using this drop-down list, you can select the valve number ( through
4) that you want to configure.
B – Valve Maximum. This field is used to set the maximum valve that can be entered for the valve gas
flow. Tap on the field to change it.
C – Valve Decimal Place. This field is used to set the number of decimal places that will be used in the
valve gas flow value in the Nitrider Calculation page. For example, if the valve decimal place is set to “1”
in this menu, and “600” is entered for valve 1 flow on the Nitrider Calculation page, the displayed value
will be “60.0”. NOTE: This field will also change the number of decimal places used for the Valve
Maximum field in the Valve Setup menu.
D – Gas Selection Drop-Down List. This drop-down list allows you to select the gas used in the selected
flow control valve. If used, valve 1 must be used with N
2
, and valve 2 must be used with NH
3
(these are
fixed). Valve 3 may be used with Dissociated Ammonia (DA), NH
3
, a miscellaneous gas (Misc),
endothermic gas (Endo), or CO
2
. Valve 4 may be used with H
2
, NH
3
, a miscellaneous gas (Misc),
endothermic gas (Endo), or CO
2
.
E – “Enabled” checkbox. If this box is checked, the selected flow control valve number is enabled and will
be displayed on the Nitrider Calculation page. If it is not checked, the valve is disabled; it will not be
displayed on the Nitrider Calculation page.
When finished with valve setup, tap the “Done” button.
Shut Down Interface
Use this option to shut down the touch screen interface for the MGA 6010. It is recommended that you not
shut down the screen interface unless you are following technical support instructions from SSi or you
are preparing to power down the MGA 6010 unit.