Gas alarm setup, Instrument information, General information – Super Systems MGA6010 User Manual
Page 37
MGA 6010 Operations Manual
Super Systems Inc.
Page 37 of 54
Gas Alarm Setup
The MGA 6010 allows the user to configure various alarms. For each parameter, the Lower Limit, Upper
Limit, and Action. As default, no alarms are enabled on the MGA unless a special request was made to do
so at the time of ordering. To modify a parameter, select the item and press Edit. A screen will appear
allow for the entry of the limits and the action. When any parameter is in an alarm state, a notification
banner will show at the top of the screen identifying the alarm, and the red light on the door will
illuminate. If desired, one of the two relays (or both simultaneously) can be energized. There are four
possible actions for the alarms:
•
None – On screen notification of alarm condition. No relay actions.
•
AL1 – On screen notification of alarm condition plus energizing of alarm relay #1.
•
AL2 – On screen notification of alarm condition plus energizing of alarm relay #2.
•
AL1 & AL2 – On screen notification of alarm condition plus energizing of alarm relays #1 and #2.
The relays provide a contact that can be connected to external lights, horns, or other devices as desired.
The rating for these relay contacts is maximum of 6 Amps / 250 Volts.
Instrument Information
These items cannot be modified; they can only be viewed.
General Information
This is information on the revision levels of various components of the instrument. This can be valuable
when consulting with Super Systems about issues with the instrument.