Seaward 300 PrimeTest User Manual
Page 82

PrimeTest 300
Operating Instructions
- 81 -
14.4
Class II Leakage Current
Test Voltage
Mains supply 230V +10% / -15%
Max Load Current
10A
Input Impedance
2KΩ As per IEC 61010-1 A.1
Test Duration
2, 5, 10, 20, 30 seconds
Pass/Fail Limit
User Defined
Display Range
0.00 – 3.50mA
Measuring Range
0.10 – 3.50mA
Resolution
0.01mA
Accuracy
± 5% ±5 digits
Attention: Avoid prolonged, repeated use at
full load and excessive test duration as this
may reduce the life of the unit.
14.5 Class
I Leakage Current
Measurement Method
Differential
Test Voltage
Mains supply 230V +10% / -15%
Max Load Current
10A
Test Duration
2, 5, 10, 20, 30 seconds
Pass/Fail Limit
User Defined
Display Range
0.00 – 19.99mA
Measuring Range
0.25 – 19.99mA
Resolution
0.01mA
Accuracy
± 5% ±5 digits
Attention: Avoid prolonged, repeated use at
full load and excessive test duration as this
may reduce the life of the unit.
14.6 Earth
Loop Resistance
Test Voltage
Mains supply 230V +10% / -15%
Test Current
Non-trip for 30mA RCD
Test Duration
Approx 10 seconds
Pass/Fail Limit
User Defined
Display Ranges
0.0 – 199Ω
Measuring Range
1.0 – 199Ω
Resolution
0.1Ω, 0.0 – 19.9Ω
1Ω,
20 – 199Ω
Accuracy
± 10% ±
1.2
1.2Ω