QuickLabel QLS-4100 Xe User Manual
Page 45

QLS-4100 Xe
3-13
Calibrating Continuous Media
1
Press the PRINT/PAUSE key until the printer is in a paused state.
2
Press the NEXT or PREVIOUS key until the Cal & Setup menu is 
displayed, then press the SELECT key. 
3
Press the NEXT or PREVIOUS key until the Calibrate menu is displayed.
4
Press the + or - key until Continuous is displayed.
5
Press the SELECT key. The printer will feed a short length of media and 
will then be calibrated for continuous media. The display will show the 
following prompt: 
6
Press the PRINT/PAUSE key to return the printer to a Ready state.
Calibrating Inverted Gap Media
Some examples of inverted gap media include clear labels with an opaque 
liner, a material which has sensing marks on the label itself, or a material on 
which you must sense on a preprinted image. 
When using media of this type, the printer is calibrated using the Gap 
Calibration method outlined previously in this chapter. 
The only difference in the procedure is that you would align the sensing 
mark when the printer prompts for placement of the liner or gap. 
Continuous 
Media Calibrated 
Calibrate 
Continuous 
Pause: Cal & Setup 
 Press SELECT 