Final Cleaning, Inspection and Cure Time
Before using the encoder for servo control, clean the tape scale, index marker and limit markers using alcohol
or acetone and a lint-free cotton cloth or swab. Finally, inspect the tape scale's surface for scratches, adhesive
spots or smears in the measuring length.
A cure time of 12 hours is required for the tape scale's pressure sensitive adhesive to achieve the best
performance and reliability.
Reworking to Correct Mistakes
Once installed, the tape scale, the index marker, and the limit markers cannot be moved or removed and
reinstalled. Reworking will require removal and discarding of the old tape or markers, and installation of
new ones.
If only the index or limit markers have to be replaced, the scale can remain in place and be reused so long as
there is no damage to the scale’s surface. Do not use a tool made of metal or other hard material to remove
the markers. Clean the scale’s surface completely of any adhesive residue before applying new markers.