beautypg.com

4 it-r specifications – K-Patents PR-03 User Manual

Page 99

background image

99

99

99

99

9 Indicating transmitter specifications

93

9.3.2 Interconnecting cable model code

PART NUMBER AND DESCRIPTION

PART NO.

PR-8300 =

Interconnecting cable between transmitter and sensor

PR-8300

PR-8040 =

Interconnecting cable (-PC sensor option only)

PR-8040

(for connection between transmitter and disconnecting unit)

Cable length
-010 =

10 meters (33 feet), standard length

-010

-_ _ _ =

Specify cable length in meters with 10 meter increments.

-_ _ _

Maximum length is 100 meters (330 feet)

9.4 IT-R Specifications

Indicating transmitter
Transmitter protection class:

Enclosure IP66, Nema 4X

Indicating transmitter weight:

4.5 kg

Display:

256 x 128 pixels graphic liquid crystal display (LCD)

Current output:

4–20 mA/0–20 mA, max. load 1000 Ohm, galvanic isolation
1500 V DC or AC (peak), built-in hold function during prism
wash

Serial output:

RS485/RS232, galvanic isolation 500 V CD or AC (peak)

Power:

100–115 V/220–240 V, 50/60 Hz, 20 VA

Alarms:

Two built-in signal relays, max 24 V, 500 mA, DC/AC

Approved environmental conditions
Ambient temperature:

max. 50

C (122

F), min. 0

C (32

F)

Humidity:

20–100 %

Overvoltage category:

Category II

Pollution degree:

Degree 2

Interconnecting cable

Shielded cable, 2 twisted pairs with individual shields, 0.5 mm

2

Digital transmission according RS485

Interconnecting cable length:

Standard 10 m (33 ft), max. 100 m (330 ft)

Cable fittings options:

European cable glands or US conduit hubs

9.4.1 Fuses

The following 250 V fuses according to IEC 60127 are printed on circuit board PR-7030:

Fuse F1, F2 : 5 x 20 mm, T1A (slow)

AC Main power protects electronics against
wrong primary voltage

Fuse F3

: 5 x 20 mm, T63 mA (slow)

4-20 mA output protection

Fuse F4

: 5 x 20 mm, T2A (slow)

Secondary main fuse

Fuse F5

: 5 x 20 mm, T500 mA (slow)

Sensor power protection

Fuse F6

: 5 x 20 mm, T500 mA (slow)

Serial bus protection

Fuse F7

: 5 x 20 mm, T1A (slow)

Processor card protection

Important: