9 Indicating transmitter specifications
97
9.7 Display messages
Message
Cause and action
NORMAL OPERATION
No faults.
DIVERT CONTROL
Not used with PR-03.
If this message occurs, deactivate with the command
Calibrate / Parameters / Divert control /
Divert Control
.
REMOVED FROM DIVERT CONTROL
Not used with PR-03
RETRYING WASH
The prism wash failed the wash check (Section 5.9.4). The
(Sanitary refractometer PR-03-A
prism wash will be attempted as many times as is set by the
and probe refractometer PR-03-P)
Wash retry
parameter (Section 5.9.4).
WASH STOP/TEMP LIMIT
The process temperature is below the temperature limit set for
(Sanitary refractometer PR-03-A
wash stop (Section 5.9.3). The message is shown until wash
and Probe refractometer PR-03-P)
succeeds or the RESET key is pushed.
LOW CONCENTRATION/NO SAMPLE
Normally caused by empty process pipe or exceptionally low
concentration.
If pipe is full and sample indicates normal concentration,
check the optical image (Section 7.2).
HIGH PROCESS TEMPERATURE
Process temperature is above the set limit (Section 5.7). Com-
mon causes are abnormal process conditions or a leaking
prism wash steam valve.
LOW LIGHT INTENSITY
Most common cause is onset of prism coating.
DIP IN IMAGE
Indicates irregular optical image. Can be caused by a speck of
dust on the CCD window.
HIGH CONC/PRISM COATED
Is most commonly caused by prism coating, but may also indi-
cate exceptionally high concentration. If sample is normal and
the message persists after the prism surface is cleaned, check
the optical image (Section 7.2).
** WASH STOP/EXT. STOP **
Caused by closing an input switch that has been activated for
(Sanitary refractometer PR-03-A
external wash stop (Section 3.2.2.6 and Section 5.6).
and probe refractometer PR-03-P)