Ab c d – K-Patents PR-03 User Manual
Page 64

64
64
64
64
58
PR-03 instruction manual
IMAGE DIAGNOSTICS
Max intensity OK (above 100)?
Endpoint below 75%?
Image below corner?
Slope OK (above 1)?
Left curve L) OK (above 1)?
Right curve R) OK (above 1)?
Yes
Yes
Yes
Yes
Yes
Yes
SENSOR HEAD
Head temperature: 28
°C
Head humidity: 3%
Normal operation.
A
B
C
D
CONC
68.0%
PROCESS TEMPERATURE: 31.2 °C
STANDARD RI(25°C): 1.4194
TEST: 115.7
Normal operation
Calibrate
Display
A
B
C
D
A
B
C
D
CONC
PROCESS TEMPERATURE: 32.5
°C
STANDARD RI(25
°C): 1.4190
TEST: 117.2
Normal operation
Optical
image
Para-
meters
68.0%
A
B
C
D
RAW
SENSOR
DATA
RMN: 2
RMX: 216
LED: 96
Scans: 2
A/D: 314
Sts: 00h
HT: 143
HH: 10
Scaled
image
Normal operation
A
B
C
D
SCALED
OPTICAL
IMAGE
TEST: 115.9
51.2
°C
Endp: 21
L(: 9.8
R): 25.9
HT: 20 °C
HH: 3%
Slope
Normal operation
Diagnos-
tics
SLOPE
TEST: 115.7
31.2
°C
Max: 208
Slope: 2.7
Endp: 21
L): 10.1
R(: 25.8
Normal operation
A
B
C
D
Good LED value 20-150
Good slope value
more than 1.5
Good HT lower than 60 ºC
Good HH lower than 10 %
Figure 7.5
Troubleshooting your refractometer system with IT-R