beautypg.com

3 disassembling sdi valve for welding – K-Patents PR-23-SD Generation 1 User Manual

Page 9

background image

Safe-Drive™ Process Refractometer PR-23-SD Best Practices, Generation 1

© Copyright K-Patents 2013. All rights reserved.

9

NOTE: Make sure that the FLOW marker is parallel to the pipe and points to the correct
flow direction. On a horizontal pipe the nozzle points downwards and on a vertical pipe
it points to right. The SD sensor system must always be installed in horizontal position
and on the side of the pipe.

3. Drill two holes – 50 mm (2”) and 25 mm (1”) – as guided by the sticker.
4. Remove the bridge between the holes so that the opening is exactly the shape of

the sticker.

Weld

Cutout

Safe Drive

mounting flange

Process pipe

Weld

Cutout

Safe Drive

mounting flange

Process pipe

Figure 6

Installation on vertical and horizontal pipes

4.3 Disassembling SDI Valve for Welding

To avoid thermal damage to the isolation valve sealing, you must separate the valve body
from the valve assembly before it is welded onto the pipe.