K-Patents PR-23-SD Generation 1 User Manual
Page 14

Document/Revision No. BP-EN-SDGEN1 1.0 Effective: April 1, 2013
14
● In addition to the SDI valve steam fittings, the following components must be included
in the steam wash installation:
● steam shut-off valve
● air-operated solenoid valve
● steam trap
● switch or terminal for power isolation
For K-Patents recommendations, please see K-Patents Steam Instructions.
● OPTIONAL, in case of contaminants: To remove any contaminants within the
steam source, installing a steam strainer is recommended. See K-Patents list of
recommended accessories.
● OPTIONAL, in case of excessive pressure: If the steam pressure exceeds to
maximum pressure differential, a pressure reducing valve (PRV) needs to be
installed to reduce the steam pressure to optimal design. See K-Patents list of
recommended accessories.
Figure 10
Mounting steam wash system
For more detailed information on steam-realted issues, please see K-Patents
Steam Instructions.
To install the wash system
1. Define the wash setting values for the wash system:
● steam source minimum and maximum pressures
● wash time – the time one wash will last (seconds)
● recovery time – the time after the wash has finished, before the measurement is
live data again (seconds)
● interval – the time between washes (minutes)