beautypg.com

K-Patents PR-23-SD Generation 1 User Manual

Page 14

background image

Document/Revision No. BP-EN-SDGEN1 1.0 Effective: April 1, 2013

14

● In addition to the SDI valve steam fittings, the following components must be included

in the steam wash installation:

● steam shut-off valve

● air-operated solenoid valve

● steam trap

● switch or terminal for power isolation
For K-Patents recommendations, please see K-Patents Steam Instructions.

● OPTIONAL, in case of contaminants: To remove any contaminants within the

steam source, installing a steam strainer is recommended. See K-Patents list of
recommended accessories.

● OPTIONAL, in case of excessive pressure: If the steam pressure exceeds to

maximum pressure differential, a pressure reducing valve (PRV) needs to be
installed to reduce the steam pressure to optimal design. See K-Patents list of
recommended accessories.

Figure 10

Mounting steam wash system

For more detailed information on steam-realted issues, please see K-Patents
Steam Instructions
.

To install the wash system

1. Define the wash setting values for the wash system:

● steam source minimum and maximum pressures

● wash time – the time one wash will last (seconds)

● recovery time – the time after the wash has finished, before the measurement is

live data again (seconds)

● interval – the time between washes (minutes)

This manual is related to the following products: