26
● Connect the cable shield to terminal 3.
● Tighten up the cable gland.
● Screw the nameplate back on.
NOTE: To avoid damage from stray voltages and short-circuiting, always disconnect
the sensor cables from the transmitter before removing the sensor.
Figure 13
Sensor electrical connections
3. Open the front panel by loosening the front panel screw.
WARNING: Always check that the power is off before opening the front panel. If the
green power indicator light is on, there is still power in the system. To completely
turn off the power, unplug the power supply cord or use the external power switch
(if installed).
Figure 14
Opening transmitter front panel
This manual is related to the following products: